accredited

accredited to(ISO/IEC 17025)

IMV TECHNO VIETNAM CO., LTD is accredited to ISO/IEC 17025:2017.

IMV TECHNO VIETNAM CO., LTD has been accredited by BOA and operating under quality control management systems in accordance with ISO / IEC 17025:2017 on 2nd March 2020 which specifies the testing capability.

Outline

Bureau of Accreditation Vietnam (BoA)

  • ① Accreditation number: VILAS 1284
  • ② Certification Bodies: Bureau of Accreditation Vietnam (BoA)
  • ③ Accreditation date: 2nd March 2020
  • ④ Accreditation scope: Vibration test (Sinusoidal), Vibration test (Broad band random), Shock test, Dry heat environmental test, Cold environmental test, Change of temperature test, Damp heat environmental test (steady), Damp heat environmental test (cyclic).

ISO/IEC 17025:2017 is the international standard which specifies “General requirements for the competence of testing and calibration laboratories”. ISO 9001 is only applicable to the quality management system; however ISO/IEC 17025 involves the requirement for both quality management and technical competence in vibration testing.
This certificate shows that the quality management system and technical capability for testing in our test laboratory is acknowledged globally and IMV proudly offers our customers test results accredited to ISO/IEC 17025.

Scope of accreditation

NO Materials or
product tested
The name of
specific tests
Limit of quantitation (if any)/
range of measurement
Test method
9

- Electrical and
electronic components

- Mechanical components

- Transportation vehicle
components

- Consumer products

- Medical devices

Vibration test
(sinusoidal)
- Frequency: ≤ 2600Hz
- Max. displacement: 76.2mmp-p
- Max. acceleration: 855m/s2o-p
IEC 60068-2-6:2007
ISO 16750-3:2012
10 Vibration test
(broad band random)
- Frequency: ≤ 2600Hz
- Maxi. displacement: 82mmp-p
- Maxi. acceleration: 598m/s2rms
IEC 60068-2-64:2008
ISO 16750-3:2012
11 Shock test - Maxi. displacement: 82mmp-p
- Maxi. acceleration: 2080m/s2o-p
IEC 60068-2-27:2008
ISO 16750-3:2012
12 Dry heat environmental test - Temperature: ≤ 180°C
-Temperature ramp rate: 1.3°C/Minute
IEC 60068-2-2:2007
ISO 16750-4:2010
13 Cold environmental test - Temperature: ≥ -70°C
- Temperature ramp rate: 1.3°C/Minute
IEC 60068-2-1:2007
ISO 16750-4:2010
14 Change of temperature test - Temperature: -70°C to +180°C
- Temperature ramp rate: 1.3°C/Minute
IEC 60068-2-14:2009
ISO 16750-4:2010
15 Damp heat,
cyclic environmental test
- Temperature: ≤ +180°C
- Temperature ramp rate: 1.3°C/Minute
- Humidity: 20%RH to 98%RH
IEC 60068-2-30:2005
ISO 16750-4:2010
16 Damp heat,
steady state environmental test
- Temperature: ≤ +180°C
- Humidity: 20%RH to 98%RH
IEC 60068-2-78:2012
ISO 16750-4:2010